MILKOSCOPE Julie C3 Automatic
Julie C3 Automatic is the successor of Julie C3.
Case – Aluminum 3 mm
LCD display, 4 rows, 16 symbols
Scope – system for automatic start of the analysis
Built-in device for coping with distortions, caused by electricity malfunctions.
System for control of the sound signal.
PC connection option with a cable and software, included in the analyzer’s kit.
Thermal printer connection option – to be chosen by the client.
The milk analyzer has 3 calibration channels, which can be set for 3 types of milk, chosen by the client.
One calibration is included in the analyzer’s price – the other/s should be paid.
Cleaning – automatic, with the MilkoScope – Flush cleaning concentrate.
Written and sound message for necessity of cleaning.
Suitable for all kinds of animal milk.
Measuring Ranges:
|
Теchnical Overview:
|
Fat: |
0% - 15%
|
Sample Volume: |
5-12 ml.
|
Protein: |
0% - 8%
|
Sample Speed: |
65 sec.
|
Lactose: |
0% - 7%
|
Sample Temperature: |
5 - 30°C
|
Solids-Non-Fat (SNF): |
0% - 15%
|
Humidity: |
0 - 80% RH
|
Solids: |
0% - 5%
|
Database Processing: |
RS 232 port
|
Density: |
1000-1200 kg/m3
|
Power Supply: |
110/230VAC±10%,50/60Hz
|
Added Water: |
0% - 100%
|
Power Consumption: |
30W
|
Freezing Point: |
-0,400 tо -0,700°C
|
Dimension: |
H=280 mm,W=170 mm,D=245 mm
|
Sample Temperature: |
5 - 30°C
|
Weight: |
3,1 kg
|
Accuracy:
|
Кit:
|
Fat: |
±0.04%
|
against Rose Gottlieb
|
Мilk Analyzer |
|
Protein: |
±0.03%
|
against Kjeldahl
|
Power Supply Cable 220V |
|
Lactose: |
±0.03%
|
against Polarimetry
|
Serial Cable to PC |
1piece
|
Solids-Non-Fat (SNF): |
±0.05%
|
against Drying Oven
|
Sample Cup |
1piece
|
Solids: |
±0.03%
|
against Gravimetric
|
Waste Container |
1piece
|
Density: |
±0,3 kg/m3
|
against pycnometer
|
Spare Hose |
1piece
|
Added Water: |
±2%
|
against cryoscope
|
Cleaning Concentrate |
1piece
|
Freezing Point: |
0,005 °C
|
|
User’s Manual |
|
|
|
|
CD with software |
|
Deviation at numerous analyses of one and
the same sample: ±0.02% for all parameters.